Parametric Rule-Based Intelligent System (PRISM) for Design and Analysis of High-Strength Separable Microneedles

A tensile test divider icon

Ju S, Im S, et al.

Micromachines

Chonnam National University, Korea Institute of Medical Microbotics, Chungbuk National University

RESEARCH SUMMARY
This study presents PRISM (Parametric Rule-Based Intelligent System), a computational platform integrating parametric 3D modeling, structural analysis, and high-resolution micro-3D printing for the design of high-strength separable microneedles. PRISM automatically generates tunable 3D geometries based on user inputs (barb number, tip angle, neck ratio, and cavity size) and predicts output metrics including surface area, volume, and cross-sectional area at the neck to optimize mechanical robustness and separability. Prototypes fabricated via projection microstereolithography (pµSL) showed sub-5 µm printing precision and <5% dimensional error from CAD models. The optimized design (Group D) with a narrowed neck, internal cavity, and triple-barb reinforcement achieved 2.13 ± 0.51 N axial strength and 73.92 ± 34.77 mN shear fracture force—outperforming conventional geometries. Skin phantom tests confirmed reliable tip detachment and retention, validating PRISM as a rapid design–testing pipeline for transdermal delivery.
CellScale hexagons, without text

CELLSCALE INSTRUMENT USED

MicroTester

Shear-direction separability of PRISM-designed microneedles was quantified using a CellScale MicroTester G2. Microneedle samples (prepared as 1 × 2 arrays with 1.5 mm spacing) were mounted in a custom cartridge holder to ensure consistent alignment and fixation. The MicroTester applied a controlled lateral shear displacement until detachment, using a constant-rate protocol over 15 s to reach a total of 2 mm displacement per detachment event. During the test, the MicroTester continuously recorded force, and the peak shear force immediately prior to detachment was extracted as the shear fracture (separation) force for each microneedle design group (n = 5 per group).
AUTHORS

Sanghwi Ju, Seung-hyun Im, Kyungsun Seo, Junhyeok Lee, Seokjae Kim, Tongil Park, Taeksu Lee, Byungjeon Kang, Jayoung Kim, Ryong Sung, Jong-Oh Park, Doyeon Bang.

PUBLICATION DETAILS
JOURNAL

Micromachines

YEAR

2025

INSTITUTIONS

Chonnam National University, Korea Institute of Medical Microbotics, Chungbuk National University

COUNTRIES

South Korea

INSTRUMENT USED

MicroTester

TESTING METHODS

Micro-Mechanical TestingShear TestingUltra Low Force Testing

RESEARCH APPLICATIONS

Drug Screening & Drug Delivery MechanicsInjectable & Regenerative BiomaterialsPolymers and Elastomers TestingSkin and Wound Healing Biomechanics

Related Publications:

Instrument Used:
Year:
Testing Method:
Research Application:
Country:

Biomechanics and composition of fixed neurocranial tessellated cartilage from the neurocranium of four mobulid ray species with different depth ranges

Hinrichs T, Zhong J, Chang CW, et al.

Journal of the Royal Society Interface

MicroTester

Hydrated and Temperature Controlled TestingIndentation TestingMicro-Mechanical Testing

Cartilage and Meniscus Mechanics

2026

FGF-2 signaling dominance in TGF-β1/FGF-2–primed nasoseptal chondrocytes for cartilage tissue engineering

Ma Z, Grimes K, et al.

Journal of Tissue Engineering

MicroTester

Compression TestingMicro-Mechanical TestingStress Relaxation TestingViscoelastic & Time-Dependent Testing

Cartilage and Meniscus MechanicsMicrotissue and Spheroid Mechanics

2026

Photodynamic treatment of glioblastoma plus endothelial cell spheroid models: increased proliferative and migratory aggressiveness of surviving tumor cells due to iNOS/NO upregulation

Fil M, Lach M, Bazak J, et al.

Photochemical & Photobiological Sciences

MicroTester

Compression TestingMicro-Mechanical Testing

Cancer MechanobiologyMicrotissue and Spheroid MechanicsOrganoid and Tissue Mimetic Systems

2026

Contact Sales

Product of Interest:
CellScale hexagon shapes