Long-term remodeling of compliance matched tissue engineered vascular grafts is sex specific

A tensile test divider icon

Martinet KM, Maestas DR Jr, Min LX, et al.

Acta Biomaterialia

University of Pittsburgh, Vascular Medicine Institute, McGowan Institute for Regenerative Medicine

RESEARCH SUMMARY
This study investigated how biological sex influences the long-term remodeling of compliance-matched tissue-engineered vascular grafts. Multilayer electrospun grafts were fabricated from polyester sulfobetaine urethane urea, polyester urethane urea, and gelatin, with the relative thicknesses of high- and low-gelatin layers adjusted to produce compliance-matched TEVGs and stiffer hypocompliant TEVGs. Benchtop mechanical characterization confirmed that compliance-matched grafts had a mean compliance of 0.0011 ± 0.0002 mmHg^-1, statistically indistinguishable from native rat abdominal aorta at 0.0013 ± 0.0004 mmHg^-1, whereas hypocompliant grafts were significantly less compliant at 0.0007 ± 0.00009 mmHg^-1. Grafts were implanted as abdominal aortic interposition grafts in male and female Sprague Dawley rats and monitored for six months. Immediately after implantation, compliance-matched grafts remained within the native aortic compliance range and were significantly more compliant than hypocompliant grafts, although their compliance decreased during the first month and subsequently became similar to the hypocompliant group. All implanted grafts remained patent through six months. Long-term remodeling depended on both graft mechanics and biological sex. Compliance-matched grafts showed greater degradation and less calcification than hypocompliant grafts. In males, compliance matching increased expression of contractile vascular smooth muscle cell genes including Acta2, Myh11, and Cnn1. In females, compliance matching reduced residual CD68-positive macrophage presence compared with hypocompliant grafts. Independent of graft type, female animals exhibited greater intimal thickening, reduced luminal mature VSMC markers, greater fibrillar collagen deposition, increased scaffold degradation, and decreased calcification compared with males. Overall, the study shows that initial mechanical compliance matching can improve several long-term TEVG remodeling outcomes even when the graft does not maintain native-like compliance, while also demonstrating that the biological response to graft mechanics is strongly sex dependent.
CellScale hexagons, without text

CELLSCALE INSTRUMENT USED

Custom CellScale Tubular Testing System

A Custom CellScale Tubular Testing System was used to measure the pressure-diameter compliance of native rat abdominal aortas and electrospun tissue-engineered vascular grafts before implantation. Native abdominal aortas from six Sprague Dawley rats, including three males and three females, were explanted and stored in sterile saline before testing. Native aortas and TEVGs were cannulated onto 1 mm outer-diameter cannulas and secured within the CellScale system. Before pressurization, each specimen was pulled taut until the system load cell measured 0.01 N, providing a standardized axial preload. Samples were then internally pressurized from 0 to 140 mmHg at a pressure ramp rate of 4 mmHg/s. Ten pressure cycles were performed, with the first nine used for preconditioning and the tenth retained for compliance analysis. Eight compliance-matched and eight hypocompliant TEVGs were tested. All mechanical testing was controlled using CellScale OpenX 12.11 software. Compliance was calculated from the change in outer diameter between 70 and 120 mmHg, normalized to the diameter at 70 mmHg and the corresponding 50 mmHg pressure interval. The custom CellScale system demonstrated that compliance-matched TEVGs had a mean in vitro compliance of 0.0011 ± 0.0002 mmHg^-1, closely matching the native aortic value of 0.0013 ± 0.0004 mmHg^-1, whereas hypocompliant grafts measured 0.0007 ± 0.00009 mmHg^-1 and were significantly less compliant than both native aorta and compliance-matched TEVGs. These measurements established the mechanical distinction between the two engineered graft designs and verified that the compliance-matched formulation achieved the intended native-like pressure-diameter response before implantation.
AUTHORS

Katarina M. Martinet, David R. Maestas Jr., Leon X. Min, Reyhaneh Gholami, Keishi Kohyama, Jacqueline L. Avila, Sang Ho Ye, Kang Kim, William R. Wagner, Jonathan P. Vande Geest.

PUBLICATION DETAILS
JOURNAL

Acta Biomaterialia

YEAR

2026

INSTITUTIONS

University of Pittsburgh, Vascular Medicine Institute, McGowan Institute for Regenerative Medicine

COUNTRIES

United States

INSTRUMENT USED

Custom CellScale Tubular Testing System

TESTING METHODS

Hydrated and Temperature Controlled TestingPressure Testing

RESEARCH APPLICATIONS

Polymers and Elastomers TestingScaffold Mechanical TestingVascular Tissue Engineering & Mechanics