RESEARCH SUMMARY
This study developed electrospun polycaprolactone/zein scaffolds containing the citrus-derived flavonoids naringin or naringenin and investigated how polymer blend composition and flavonoid identity affect mechanical properties, drug release, water uptake, degradation, antioxidant and antibacterial activity, and preliminary osteogenic responses. PCL and PCL/zein scaffolds containing 10 or 20 wt.% zein were fabricated with or without 5 wt.% naringin or naringenin. All formulations produced homogeneous interconnected electrospun fibers, with increasing zein content reducing average fiber diameter from approximately 1.53 µm for neat PCL to 0.53 ± 0.20 µm for PCL containing 20% zein. Mechanical properties depended strongly on the interaction among PCL, zein, and the incorporated flavonoid. PZ10 exhibited a Young’s modulus of 8.78 ± 0.86 MPa and tensile strength of 3.02 ± 0.10 MPa, whereas increasing zein to 20% without flavonoid reduced stiffness, strength, and elongation. Naringin and naringenin altered these trends differently: PZ20NI reached a modulus of 10.8 ± 1.0 MPa, while PZ20NE exhibited the highest modulus of all formulations at 15.22 ± 1.42 MPa with a tensile strength of 2.75 ± 0.11 MPa. Polymer composition also controlled hydration and therapeutic release. Increasing zein raised water uptake, while naringin- and naringenin-loaded fibers sustained flavonoid release over 72 hours. PZ20NE showed the strongest antioxidant response at 49.6% DPPH inhibition and nearly complete inhibition of Staphylococcus aureus, together with approximately 85% inhibition of Escherichia coli. All scaffold extracts maintained MG-63 osteoblast-like cell viability at or above approximately 79%, and zein- and flavonoid-containing formulations increased alkaline phosphatase activity relative to neat PCL. PZ20NI produced the strongest ALP response among the tested materials. Overall, the results show that PCL/zein ratio and flavonoid structure can be used to tune the mechanical, drug-release, antimicrobial, antioxidant, and osteogenic characteristics of electrospun scaffolds, supporting further development for bone tissue repair.
CellScale hexagons, without text

CELLSCALE INSTRUMENT USED

BioTester

A CellScale BioTester 3000 equipped with a 10 N load cell was used for uniaxial tensile testing of electrospun PCL, PCL/zein, and flavonoid-loaded PCL/zein scaffold specimens. Five rectangular specimens were randomly cut from each scaffold formulation and tested individually at a displacement rate of 0.5 mm/min. The protocol followed ASTM D882-12 for tensile testing of thin plastic films less than 1 mm thick. BioTester force and displacement measurements were converted to engineering stress and engineering strain using the initial cross-sectional area and gauge length of each specimen. Stress-strain curves were then used to calculate elastic modulus from the linear elastic region, ultimate tensile strength as the maximum measured stress, and elongation at break from the change in specimen length at failure. The BioTester results demonstrated that scaffold mechanics depended on both zein concentration and flavonoid identity. PZ10 reached 8.78 ± 0.86 MPa Young’s modulus and 3.02 ± 0.10 MPa tensile strength, while PZ20 showed reduced stiffness and strength. Adding naringin to the 20% zein formulation increased modulus to 10.8 ± 1.0 MPa and tensile strength to 1.97 ± 0.21 MPa, whereas naringenin produced the strongest stiffening response, with PZ20NE reaching the highest Young’s modulus of 15.22 ± 1.42 MPa and a tensile strength of 2.75 ± 0.11 MPa. The BioTester measurements therefore provided the quantitative mechanical evidence that polymer ratio and flavonoid-polymer interactions can substantially alter electrospun scaffold stiffness, strength, and extensibility, helping identify mechanically promising formulations for future bone-regeneration applications.
AUTHORS

Cinthya Enriquez-Lara, Jonathan Manríquez-Gallardo, Esmeralda Lopez, Fabián Álvarez-Carrasco, Daniel Canales, Alejandro Vásquez-Espinal, Raúl Rangel-López, Juan José Martínez Sanmiguel, Diana G. Zarate-Triviño, Paula A. Zapata, Viviana Moreno-Serna.

PUBLICATION DETAILS
JOURNAL

Journal of Functional Biomaterials

YEAR

2026

INSTITUTIONS

Universidad Arturo Prat, Universidad de Santiago de Chile, Universidad de Las Américas, Universidad Autónoma de Nuevo León, Centro de Ingeniería y Desarrollo Industrial

COUNTRIES

Chile, Mexico

INSTRUMENT USED

BioTester

TESTING METHODS

Tensile Testing

RESEARCH APPLICATIONS

Bone Tissue Engineering & MechanicsDrug Screening & Drug Delivery MechanicsMembranes and Thin Films MechanicsPolymers and Elastomers TestingScaffold Mechanical Testing

Related Publications:

Instrument Used:
Year:
Testing Method:
Research Application:
Country:

Effect of electron beam irradiation on the mechanical and solubility properties of nanofibrous electrospun polyurethane composite biomaterials

Webb CWB, Chen Y, D'Costa K, Santerre JP

Materials Letters

BioTester

Tensile Testing

Membranes and Thin Films MechanicsPolymers and Elastomers TestingScaffold Mechanical Testing

2027

Multiscale Characterization by Equibiaxial Tensile Testing and Time-Domain 1H Solid State NMR for PTMC-Based Networks for Tissue Growth

Feyne F, van Bochove B, Spoljaric S, et al.

Macromolecular Materials and Engineering

BioTester

Biaxial TestingViscoelastic & Time-Dependent Testing

Polymers and Elastomers Testing

2026

Electrospun PCL/Zein Fibers with Naringin or Naringenin for Potential Bone Repair Applications: Physicochemical and Preliminary Biological Evaluation

Enriquez-Lara C, Manríquez-Gallardo J, Lopez E, et al.

Journal of Functional Biomaterials

BioTester

Tensile Testing

Bone Tissue Engineering & MechanicsDrug Screening & Drug Delivery MechanicsMembranes and Thin Films MechanicsPolymers and Elastomers TestingScaffold Mechanical Testing

2026

Contact Sales

Product of Interest:
CellScale hexagon shapes