Comparative Analysis of Two Measurement Modalities for Ex Vivo Analysis of Corneal Stiffness in Porcine Corneas

Reifeltshammer SA, Seferovic H, et al.

Bioengineering

Johannes Kepler University Linz, Kepler University Clinic, University of Bern, Ludwig - Maximilians - University Munich, IROC Zurich

RESEARCH SUMMARY
This study compared two widely used biomechanical testing modalities—uniaxial tensile testing and AFM nanoindentation—to characterize stiffness changes in porcine corneas following corneal crosslinking (CXL). Sixty-two eyes were assigned to control, Dresden-protocol CXL (CXL3), or accelerated CXL (CXL9) groups, then evaluated using both modalities on paired corneal samples. Uniaxial tensile testing demonstrated a clear stiffening response, with CXL3 producing the largest increase (≈1.96-fold increase in Young’s modulus at 10% strain), followed by CXL9 (≈1.48-fold). AFM nanoindentation detected similar directional trends, with localized stiffness increases of 16–28%, though group differences were not statistically significant due to spatial heterogeneity at the micrometer scale. The findings highlight how tensile testing captures global stiffening across the entire tissue strip, while nanoindentation reveals position-dependent stromal variations. Together, the modalities offer complementary insights into the biomechanical effects of different CXL protocols.

CELLSCALE INSTRUMENT USED

UStretch

Corneal strips were mounted on a CellScale UStretch uniaxial tensile testing device equipped with a 1 N load cell. Samples were clamped 4 mm apart and immersed in 16% dextran to maintain hydration. Each specimen underwent a 120 s preload at 5×10³ Pa followed by controlled extension at 0.035 mm/s to 16% strain. The UStretch recorded force–displacement data at 10 ms intervals, enabling calculation of stress–strain curves and derivation of Young’s modulus at 6%, 8%, and 10% strain. These measurements provided the global stiffness comparisons between control, CXL3, and CXL9 tissues.
AUTHORS

Sophia A. Reifeltshammer, Hannah Seferovic, Malavika H. Nambiar, Philippe Büchler, Theo G. Seiler, Jascha Wendelstein, Matthias Bolz, Peter Hinterdorfer, Yoo Jin Oh, Isaak Fischinger.

PUBLICATION DETAILS
JOURNAL

Bioengineering

YEAR

2025

INSTITUTIONS

Johannes Kepler University Linz, Kepler University Clinic, University of Bern, Ludwig - Maximilians - University Munich, IROC Zurich

COUNTRIES

Austria, Germany, Switzerland

INSTRUMENT USED

UStretch

TESTING METHODS

Hydrated and Temperature Controlled TestingTensile Testing

RESEARCH APPLICATIONS

Ophthalmic Biomechanics & Corneal Tissue Engineering

Related Publications:

Instrument Used:
Year:
Testing Method:
Research Application:
Country:

The Biomechanical Properties of Porcine Intervertebral Disc Tissue Treated with Different Fixation Solutions

Chow N, Sinopoli SI, et al.

Journal of the Mechanical Behavior of Biomedical Materials

BioTesterUStretch

Peel TestingTensile Testing

Intervertebral Disc Biomechanics

2026

Tissue-adhesive hydrogel–MXene biosensor for in situ intraoral TNF-α detection

Wong TH, Liu W, et al.

Science Advances

UStretch

Hydrated and Temperature Controlled TestingPeel TestingTensile Testing

Dental & Oral Tissue BiomechanicsHydrogel Mechanical TestingWearable Bioelectronics

2026

An Investigation of the Mechanism of Adjacent Segment Disease in a Porcine Spine Model

Chow N, Sinopoli SI, et al.

Clinical Biomechanics

BioTesterUStretch

Hydrated and Temperature Controlled TestingMicro-Mechanical TestingPeel TestingTensile Testing

Fibrosis & Tissue RemodelingIntervertebral Disc BiomechanicsMechanotransductionMusculoskeletal Tissue Engineering & Mechanics

2025

Contact Sales

Product of Interest: