ASTM F88/F88M
Seal Strength Testing

A tensile test divider icon

ASTM F88/F88M provides a tensile test framework for measuring the force required to separate seals in flexible barrier materials. Seal strength testing is used with packaging materials in medical, pharmaceutical, food, and consumer applications where consistent seal performance is important.

CellScale evaluated ASTM F88/F88M-based seal strength testing using a UniVert S with a 2 N load cell and self-sealing sterilization pouch material. The study examined average peel force, maximum force, repeatability, load cell selection, and force-displacement behaviour during progressive seal separation.

At A Glance

Close up of a peel test on the UniVert for ASTM F88 seal strength testing
The CellScale hexagons from the logo without text

What is ASTM F88/F88M?

ASTM F88/F88M is a seal strength test method for flexible barrier materials. The test measures the force required to separate a sealed specimen using tensile loading and can be used to evaluate average peel force, maximum force, and specimen failure behaviour.

Seal strength measurements can be useful when evaluating opening force, package performance, process validation, process control, and the consistency of a sealing process. Because the measured response can be affected by specimen geometry, material behaviour, and test configuration, these factors should remain consistent when results are being compared.

The CellScale hexagons from the logo without text

What Does ASTM F88 Seal Strength Testing Measure?​

Seal strength is not represented by a single feature of the force curve. Depending on the test objective, several parts of the seal-separation response can be useful.

Average peel force describes the representative force required to continue separating the seal over a selected region of the test. In the CellScale study, the average was calculated from the central 80% of the peel response.

Maximum force is the highest tensile load recorded as the specimen is progressively separated.

The force-displacement response can also show how separation develops through the specimen. Material bending, elongation, tearing, delamination, and seal geometry can all contribute to the measured response.

For this reason, ASTM F88 seal strength results should be interpreted together with the specimen configuration and test conditions rather than as an isolated force value.

The CellScale hexagons from the logo without text

ASTM F88 Test Techniques

ASTM F88/F88M-15 describes three tail-holding techniques for seal strength testing. The selected technique can affect the measured force, so the same approach should be maintained throughout a comparative test series.

Technique A: Unsupported

Technique A: Unsupported is when both specimen tails are secured in opposing grips while the seal remains unsupported.

Technique B: Supported 90°

Technique B: Supported 90° is when the seal is supported at approximately 90° to the specimen tails during separation.

Technique C: Supported 180°

Technique C: Supported 180° is when one specimen tail is supported against an alignment plate while the other is folded 180° over the seal.

Different tail-holding arrangements change how bending and other specimen effects contribute to the measured force. Results obtained using different techniques should therefore not be treated as directly interchangeable without considering the test configuration.

The CellScale hexagons from the logo without text

Equipment for ASTM F88 Seal Strength Testing

ASTM F88 seal strength testing requires a controlled tensile separation setup capable of recording tensile force and grip separation. For the ASTM F88/F88M-15 conditions evaluated in this study, the test system required adjustable constant-rate jaw separation, suitable grips, and force and displacement measurement.

Mechanical Test System

The test frame must provide controlled tensile motion over the required separation distance and speed. In our case, the UniVert.

A UniVert S setup for tensile testing horizontally
A close up of a medium force load cell for the UniVert tensile testing machine

Load Cell

The load cell should provide enough capacity for the expected maximum seal force while still giving useful measurement performance within the force range of interest.

Grips

The specimen must be held securely during separation. The gripping arrangement should minimize slippage and avoid unnecessary concentration of stress at the grip.

A set of grips for the UniVert for low force tensile testing

Force and Displacement Measurement

Both tensile force and grip separation are relevant measurements. The required accuracy should be confirmed against the specific revision of the standard being followed.

Selecting the highest-capacity sensor available is not necessarily the best approach for low-force seal strength testing. The measurement range should be matched to the forces generated by the specimen.

ASTM F88 Testing with the CellScale UniVert

CellScale evaluated whether the UniVert S could characterize the low-force peeling response of a commercially available self-sealing sterilization pouch using test conditions based on ASTM F88/F88M-15.

The study focused on five questions:

  • Could average peel force be measured consistently?
  • Could maximum force be identified across repeated specimens?
  • How repeatable would the response be across 12 tests?
  • Was a 2 N load cell appropriate for the low forces generated by the specimens?
  • What would the force-displacement response look like during progressive seal separation?
The CellScale hexagons from the logo without text

ASTM F88 Test Configuration Used in the CellScale Study

A tensile test divider icon

The same UniVert configuration and test protocol were used throughout the 12-specimen test series.

Parameter Test Configuration
Mechanical test system CellScale UniVert S
Test mode Tension
Load cell 2 N
Load cell accuracy ±0.2% of rated capacity
Tail-holding technique Technique A, unsupported
Grip separation rate 4.5 mm/s, equivalent to 270 mm/min
Initial grip separation 25 mm
Motion accuracy ±0.05% of desired distance
Position accuracy ±0.02 mm
Data acquisition rate 50 Hz
Number of specimens 12

The 2 N load cell has an accuracy of ±0.2% of rated capacity, corresponding to ±0.004 N. The UniVert actuator motion accuracy is within ±0.05% of the desired distance, with positional accuracy specified as ±0.02 mm. The white paper notes that ASTM F88/F88M-15 requires both tensile-force and grip-separation measurement systems to be accurate within ±2%.

The 270 mm/min grip separation rate used in the study falls within the 200 to 300 mm/min range described for ASTM F88/F88M-15.

The CellScale hexagons from the logo without text

ASTM F88 Specimen Preparation and Test Method

Self-sealing sterilization pouches were selected as the flexible barrier material for the study. Twelve specimens were prepared, each consisting of a plastic-film tail and a waxed-paper tail joined by the pouch seal.

Parameter Value
Material Self-sealing sterilization pouch
Specimen width 15 mm
Specimen length 110 mm
Conditioning time 72 h
Temperature Approximately 23°C
Relative humidity 45%
Upper grip material Plastic film
Lower grip material Waxed paper
Initial grip separation 25 mm
An overview of the UniVert S setup for ASTM F88 standard testing

Each specimen was mounted using the unsupported Technique A configuration. The plastic-film tail was secured in the upper grip and the paper tail in the lower grip. The moving grip then separated the specimen at 4.5 mm/s, equivalent to 270 mm/min.

Application-Specific Specimen Orientation

The specimen orientation used in the CellScale study is an important distinction from the conventional ASTM F88/F88M-15 configuration.

ASTM F88/F88M-15 specifies that the seal line should normally be perpendicular to the direction of pull. In this study, the sterilization-pouch specimens were intentionally tested parallel to the seal.

Why was the orientation changed?

The pouch contained multiple sealed regions. When specimens were pulled across the seal, those regions produced several distinct force responses. Testing parallel to the seal produced a more continuous peeling response and better represented the opening behaviour being investigated.

This orientation was therefore treated as an application-specific modification rather than presented as the conventional ASTM F88 specimen configuration.

That distinction matters when interpreting the results and when comparing them with tests performed using the standard orientation.

The CellScale hexagons from the logo without text

ASTM F88 Force-Displacement Analysis

Each specimen produced a force-displacement curve showing how much force was required as the pouch seal progressively separated.

Average Peel Force

Average peel force was calculated from the central 80% of the peel response.

Maximum Force

The highest tensile load recorded during each test was reported as maximum force.

Data Acquisition

Force and displacement were collected at 50 Hz, producing a detailed force-displacement record throughout the peeling process.

A force vs displacement graph from an ASTM F88 specimen test showing beginning to failure with central 80% annotation

The initial rise from zero force and the final portion of the curve were excluded because these regions do not represent the steady seal-separation response. The expanded study notes that ASTM F88/F88M-15 also describes excluding the initial loading region and the return toward zero from the average seal-strength calculation.

ASTM F88 Seal Strength Test Results

Twelve specimens were tested using the same UniVert configuration and test protocol.

0.705 N Mean Average Peel Force

1.186 N Mean Maximum Force

Result Average Peel Force Maximum Force
Mean 0.705 N 1.186 N
Range 0.608 to 0.801 N 0.980 to 1.346 N
Standard Deviation 0.0537 N 0.106 N

Across the test series, the coefficient of variation was approximately 7.6% for average peel force and 8.9% for maximum force. The mean maximum force normalized by the 15 mm specimen width was 79.1 N/m.

The results showed a relatively narrow distribution across the 12 specimens while still preserving measurable specimen-to-specimen variation.

Results table for 12 specimens undergoing seal strength testing on the UniVert S
The CellScale hexagons from the logo without text

Load Cell Selection for Low-Force Seal Strength Testing

A tensile test divider icon

Load cell selection had a noticeable effect on the quality of the measured force response.

Initial trials used a higher-capacity 10 N load cell. At the low forces produced by the sterilization-pouch specimens, the recorded response showed greater apparent signal noise. Repeating the test with the 2 N load cell produced a cleaner signal and more clearly resolved the seal-peeling behaviour.

The practical lesson is straightforward: load cell capacity should be high enough to accommodate the expected maximum load, but capacity alone is not the only consideration. For low-force seal strength testing, the sensor should also provide useful measurement performance in the actual force range being studied.

Across-seal trial curve for force vs displacement with 10N load cell
Across-seal trial curve for force vs displacement with 10 N load cell
A force vs displacement chart for a seal strength test on the UniVert
Along-seal curve for force vs displacement with 2 N load cell
The CellScale hexagons from the logo without text

Factors That Influence Seal Strength Measurements

Measured seal strength can be influenced by more than the seal itself.

The ASTM F88/F88M discussion in the expanded study identifies factors including:

Grip configuration, separation rate, and the analysis method can also affect comparisons between tests.

For comparative seal strength testing, these conditions should therefore be kept consistent across the test series wherever possible.

This is also why results obtained using different ASTM F88 techniques or different specimen orientations should be interpreted with care.

What the CellScale ASTM F88 Study Demonstrated

The UniVert S equipped with a 2 N load cell successfully characterized the low-force peel response of the self-sealing sterilization pouch material using test conditions based on ASTM F88/F88M-15.

The study demonstrated:

Across the 12 specimens, mean average peel force was 0.705 N and mean maximum force was 1.186 N. ASTM F88 White Paper (Expanded)

The CellScale hexagons from the logo without text

ASTM F88 Standards Statement

This study demonstrates seal strength testing using a UniVert configuration and test conditions based on ASTM F88/F88M-15. An application-specific specimen orientation was used. The study should therefore not be interpreted as a demonstration of full procedural conformance to ASTM F88/F88M-15.

A mechanical testing system meeting relevant force, displacement, or motion requirements does not by itself establish full procedural conformance. Specimen preparation, orientation, conditioning, fixture configuration, test parameters, analysis, reporting, and the specific revision of the standard may all affect conformance.

Users performing formal ASTM F88 testing should consult the current official standard and confirm the requirements that apply to their material, laboratory, product, or regulatory context.

The CellScale hexagons from the logo without text

ASTM F88 Seal Strength Testing Resources

A tensile test divider icon

For a shorter overview or a more detailed review of the CellScale study, both versions are available.

The CellScale hexagons from the logo without text

ASTM F88 FAQs

A tensile test divider icon

ASTM F88/F88M is a test method used to measure seal strength in flexible barrier materials. The method uses tensile loading to separate a sealed specimen while force is recorded.

ASTM F88 seal strength testing can be used to evaluate the force required to separate a seal, including average seal or peel force, maximum force, and the behaviour of the specimen during separation.

ASTM F88 applies to flexible barrier materials. In the CellScale study, the test specimens were prepared from commercially available self-sealing sterilization pouches with plastic-film and waxed-paper layers joined by a seal.

Technique A leaves the seal unsupported while the specimen tails are held in opposing grips. Technique B supports the seal at approximately 90°. Technique C supports one tail against an alignment plate while the other is folded 180° over the seal. The selected technique can influence the measured force.

For ASTM F88/F88M-15, the CellScale study references a jaw-separation range of approximately 200 to 300 mm/min. The UniVert test was performed at 270 mm/min, or 4.5 mm/s.

The test requires a controlled tensile separation system capable of measuring tensile force and grip separation, together with suitable grips for the flexible specimen. The appropriate load cell depends on the force range expected from the seal.

Average peel force was calculated from the central 80% of the force-displacement response. The initial loading region and final return toward zero were excluded from the calculation.

The CellScale study demonstrated ASTM F88/F88M-15-based seal strength testing using a UniVert S with a 2 N load cell, an unsupported Technique A configuration, and a separation rate of 270 mm/min. Because an application-specific specimen orientation was used, the study is not presented as a demonstration of full procedural conformance.

Seal strength describes the force required to separate a seal and is one useful measure when evaluating packaging performance. It does not by itself describe every aspect of package integrity. Other package or barrier characteristics may need to be evaluated separately depending on the application.

Need to Perform Seal Strength Testing?

If you are evaluating flexible barrier materials, packaging seals, or another low-force peel application, CellScale can help assess the expected force range, specimen geometry, grips, and test configuration.

The UniVert supports interchangeable load cells and fixtures for tensile, peel, and other mechanical testing applications involving flexible packaging, films, adhesives, biomaterials, and other low-force specimens.