Validation of a Modified Esophageal pH Probe for Monitoring Colon Health

A tensile test divider icon

Hill J, Miller E, et al.

Proceedings of the Design of Medical Devices Conference (DMD2025)

Brigham Young University

RESEARCH SUMMARY
This study presents and validates a modified Medtronic Bravo® pH capsule repurposed for extended in-situ monitoring of the gastrointestinal microbiome. The research addresses limitations of existing gut monitoring methods, which often rely on invasive sampling or single-point data collection. The modified capsule, designed for attachment to the cecal wall, enables continuous pH measurement over a 14-day period to assess short-chain fatty acid (SCFA) activity related to gut microbial health. Key innovations include extended battery life, epoxy encapsulation for durability, and a new suture-based anchoring system. Validation involved accelerated tensile fatigue testing and biofouling experiments in simulated cecal fluid, demonstrating that the capsule maintains accuracy and structural integrity under physiological conditions.
CellScale hexagons, without text

CELLSCALE INSTRUMENT USED

BioTester

An accelerated tensile fatigue protocol was performed using a CellScale BioTester 5000 biaxial test machine to verify the structural integrity of a modified Medtronic Bravo® pH capsule under simulated in-vivo peristaltic loading. The capsule was mounted between the BioTester terminals using suture thread: a binder clip was attached to the probe end and tied to one terminal, while the opposite end was tied through a drilled suture hole to the other terminal. The BioTester applied cyclic tensile loading to approximate cecal peristalsis over a 14-day study period (targeting ~120,960 cycles), accelerated to one cycle every ~2 seconds and including a 100 mN preload, with peak loading intended to represent peristaltic forces up to ~520 mN. Following completion of cycling, the capsule’s epoxy seam (original-to-new epoxy interface) was visually inspected under magnification for cracking, separation, or deformation, and capsule pH accuracy was checked before/after mechanical cycling using buffer solutions (pH 4, 7, 10.01).
AUTHORS

James Hill, Euan Miller, Benjamin Terry.

PUBLICATION DETAILS
JOURNAL

Proceedings of the Design of Medical Devices Conference (DMD2025)

YEAR

2025

INSTITUTIONS

Brigham Young University

COUNTRIES

United States

INSTRUMENT USED

BioTester

TESTING METHODS

Fatigue TestingTensile Testing

RESEARCH APPLICATIONS

Drug Screening & Drug Delivery MechanicsGastrointestinal and Urinary Tract BiomechanicsMaterial Fatigue and Durability

Related Publications:

Instrument Used:
Year:
Testing Method:
Research Application:
Country:

Biomechanical Remodeling of Ascending Aortic Aneurysms Across the 5-cm Surgical Threshold

Ghadie N, Labrosse M, Boodhwani M

JTCVS Open

BioTester

Biaxial Testing

Vascular Tissue Engineering & Mechanics

2026

Inflammation-Responsive baicalin-delivering glycocalyx-mimetic hydrogel-functionalized bioprosthetic valve for antithrombosis and immunomodulation

Chen L, Feng Y, Huang X, et al.

Acta Biomaterialia

BioTester

Tensile Testing

Drug Screening & Drug Delivery MechanicsECM & Decellularized Matrix MechanicsHeart Valve Tissue Engineering & MechanicsHydrogel Mechanical TestingStimuli Responsive Hydrogels Characterization

2026

Development of Large-Diameter Tissue-Engineered Vascular Grafts as an Alternative to Synthetic Conduits for Emergency Large-Vessel Trauma Reconstruction: A Preliminary Study

Lăpădatu BI, Ciucanu CC, Mureșan A, et al.

Journal of Cardiovascular Emergencies

BioTester

Tensile Testing

ECM & Decellularized Matrix MechanicsScaffold Mechanical TestingVascular Tissue Engineering & Mechanics

2026

Contact Sales

Product of Interest:
CellScale hexagon shapes